What Is a Bowie-Dick Test? A Sterile Processing Guide

Learn what a Bowie-Dick test checks, when it is used in a dynamic-air-removal steam sterilizer, and how it differs from biological monitoring.

Houcine Boutam

Houcine Boutam

15+ Years in Sterile Processing & Material Management · Pursuing BS Healthcare Administration

What Does a Bowie-Dick Test Check?

A Bowie-Dick test is an air-removal test for a dynamic-air-removal steam sterilizer, such as a prevacuum sterilizer. It helps detect problems with air removal and steam penetration. The CDC defines it as a diagnostic test of air removal, not a test that proves individual items are sterile.

For the exam, remember its purpose: air removal and steam penetration. It is one part of a monitoring system, not a replacement for physical, chemical, or biological monitoring.

When Is It Used?

CDC guidance states that an air-removal test is performed daily in an empty dynamic-air-removal sterilizer. Its infection prevention checklist specifies each day that type of sterilizer is used. Follow the sterilizer and test-pack manufacturers' instructions, applicable standards, and your facility's policies for the exact procedure, timing, placement, interpretation, and documentation.

The test applies to dynamic-air-removal steam equipment. Do not assume it is the routine air-removal test for every sterilizer or every sterilization method. For other monitoring concepts, see biological vs. chemical indicators.

How Does the Test Work in General?

The test uses a specified test pack or device containing a chemical indicator designed to challenge air removal and steam penetration. After the designated cycle, the indicator is read against the test manufacturer's interpretation instructions. A nonuniform or otherwise unacceptable result can indicate a problem; exact appearance and causes depend on the product and equipment.

For actual work, do not choose cycle settings or declare a result acceptable from a memorized number or an image in a study guide. Use the instructions for the specific test pack and sterilizer, along with facility policy.

What If the Result Fails?

Treat an unacceptable result as a reason to stop and follow the facility's escalation procedure. Notify the responsible supervisor, document the result, and follow the sterilizer and test manufacturer's instructions and department policy for removing equipment from service, investigation, retesting, and return to use. Do not process loads as normal until the responsible personnel determine the problem is resolved and the equipment is cleared under policy.

Possible causes require investigation; a single visual result does not identify the cause by itself. Depending on the circumstances, staff may need to review the test setup, equipment, and supporting records.

Bowie-Dick Test vs. Biological and Chemical Indicators

Monitoring toolWhat it helps assessKey distinction
Bowie-Dick testAir removal and steam penetration in a dynamic-air-removal steam sterilizerA specific air-removal challenge; not proof that a load is sterile.
Biological indicatorWhether resistant test microorganisms were inactivated under the tested processUsed in a defined monitoring program; it does not replace an air-removal test.
Chemical indicatorA visible response to specified process conditionsThe type and intended use matter; a color change alone does not prove sterility.
Physical monitoringCycle readings such as time, temperature, and pressureRead together with other monitoring and facility procedures.

Each measure answers a different question. For a fuller comparison, use our biological vs. chemical indicators guide.

Practice Questions

1. What is the main purpose of a Bowie-Dick test?

Answer: To evaluate air removal and steam penetration in a dynamic-air-removal steam sterilizer. It is not a biological indicator test and does not prove a particular instrument is sterile.

2. Does a passing Bowie-Dick test replace routine biological monitoring?

Answer: No. The tests assess different parts of the process. Continue all monitoring required by the device IFUs, standards, and facility policy.

3. A test result is unacceptable. What should happen next?

Answer: Follow the facility's procedure for stopping use, notification, documentation, investigation, and clearance. Do not ignore the result because a previous cycle passed.